Channelpedia

PubMed 12733040


Referenced in: none

Automatically associated channels: Kv10.1



Title: Optical photothermal detection in HPLC.

Authors: M A Proskurnin, A Yu Luk'yanov, S N Bendrysheva, A A Bendryshev, A V Pirogov, O A Shpigun

Journal, date & volume: , 2003 Apr , 375, 1204-11

PubMed link: http://www.ncbi.nlm.nih.gov/pubmed/12733040


Abstract
A mode-mismatched parallel dual-beam thermal lens spectrometer with a far-field single-channel detector system was used as a detector in HPLC. An expert estimation of the measurement results was applied to optimize the optical-scheme configuration of the spectrometer to achieve the longest linear calibration range and highest repeatability under chromatographic flow conditions. Chelates with 4-(2-pyridylazo)resorcinol were separated and determined with the limits of detection of n x 10(-8)- n x 10(-7) mol L(-1); the relative standard deviation of measurements was 46%. Xylenol Orange, 4-(2-thiazolylazo)resorcinol, and dithizone were studied as post-column reagents in thermal lens detection in ion chromatography. The limits of detection were n x 10(-8)- n x 10(-7) mol L(-1); the linear calibration ranges were about three orders; the relative standard deviation of measurements was 3-7%. A combined photothermal-refractometric detector for HPLC based on a polarization interferometer is proposed. Metal complexes as 4-(pyridylazo)resorcinol chelates (limits of detection of n x 10(-8)- n x 10(-7) mol L(-1)) and sugars (limits of detection of 10-20 ng L(-1)) were investigated as model substances. Obtained results were compared with results for traditional detectors, which show that photothermal detection has higher sensitivity than photometric and other absorption detectors.