Title: Optimizing Electrospray Interfaces Using Slowly Diverging Conical Duct (ConDuct) Electrodes.
Authors: Andrew N Krutchinsky, Julio C Padovan, Herbert Cohen, Brian T Chait
Journal, date & volume: J. Am. Soc. Mass Spectrom., 2015 Apr , 26, 659-67
PubMed link: http://www.ncbi.nlm.nih.gov/pubmed/25667060
Abstract
We demonstrate that the efficiency of ion transmission from atmosphere to vacuum through stainless steel electrodes that contain slowly divergent conical duct (ConDuct) channels can be close to 100%. Here, we explore the properties of 2.5-cm-long electrodes with angles of divergence of 0°, 1°, 2°, 3°, 5°, 8°, 13°, and 21°, respectively. The ion transmission efficiency was observed to jump from 10-20% for the 0° (straight) channels to 90-95% for channels with an angle of divergence as small as 1°. Furthermore, the 2-3° ConDuct electrodes produced extraordinarily low divergence ion beams that propagated in a laser-like fashion over long distances in vacuum. To take advantage of these newly discovered properties, we constructed a novel atmosphere-to-vacuum ion interface utilizing a 2° ConDuct as an inlet electrode and compared its ion transmission efficiency with that of the interface used in the commercial (Thermo Fisher Scientific, San Jose, CA, USA) Velos Orbitrap and Q Exactive mass spectrometers. We observed that the ConDuct interface transmitted up to 17 times more ions than the commercial reference interface and also yielded improved signal-to-noise mass spectra of peptides. We infer from these results that the performance of many current atmosphere-to-vacuum interfaces utilizing metal capillaries can be substantially improved by replacing them with 1° or 2° metal ConDuct electrodes, which should preserve the convenience of supplying ion desolvation energy by heating the electrode while greatly increasing the efficiency of ion transmission into the mass spectrometer.