Channelpedia

PubMed 23853294


Referenced in: none

Automatically associated channels: Kir2.3



Title: A 100-channel hermetically sealed implantable device for chronic wireless neurosensing applications.

Authors: Ming Yin, D A Borton, J Aceros, W R Patterson, A V Nurmikko

Journal, date & volume: IEEE Trans Biomed Circuits Syst, 2013 Apr , 7, 115-28

PubMed link: http://www.ncbi.nlm.nih.gov/pubmed/23853294


Abstract
A 100-channel fully implantable wireless broadband neural recording system was developed. It features 100 parallel broadband (0.1 Hz-7.8 kHz) neural recording channels, a medical grade 200 mAh Li-ion battery recharged inductively at 150 kHz , and data telemetry using 3.2 GHz to 3.8 GHz FSK modulated wireless link for 48 Mbps Manchester encoded data. All active electronics are hermetically sealed in a titanium enclosure with a sapphire window for electromagnetic transparency. A custom, high-density configuration of 100 individual hermetic feedthrough pins enable connection to an intracortical neural recording microelectrode array. A 100 MHz bandwidth custom receiver was built to remotely receive the FSK signal and achieved -77.7 dBm sensitivity with 10(-8) BER at 48 Mbps data rate. ESD testing on all the electronic inputs and outputs has proven that the implantable device satisfies the HBM Class-1B ESD Standard. In addition, the evaluation of the worst-case charge density delivered to the tissue from each I/O pin verifies the patient safety of the device in the event of failure. Finally, the functionality and reliability of the complete device has been tested on-bench and further validated chronically in ongoing freely moving swine and monkey animal trials for more than one year to date.